The Department of Health and Human Services is procuring a single-axis linear motion calibration system, commonly known as a shake table, along with control software for the FDA's laboratory calibration and testing needs. The system must meet specific performance characteristics, including programmable motion control, safety features, and compliance with accessibility standards. Delivery is anticipated within approximately 90 days after award to the FDA White Oak campus in Silver Spring, Maryland.
Get your fit score, win probability, and required capabilities.
The U.S.
Food and Drug Administration requires one (1) complete, turnkey single-axis (one degree-of-freedom) linear motion calibration system, commonly described as a shake table, with control software, for laboratory calibration, verification, and validation testing of inertial measurement devices.
The system shall be fully assembled, factory-tested, and factory-calibrated prior to shipment, requiring only unpacking, cable connection, and software installation by the end user.
The Government seeks sources able to provide a system meeting, at minimum, the following salient performance characteristics.
These are stated in approximate terms for market research; the full minimum specification will accompany any resulting solicitation. • Programmable single-axis linear motion under closed-loop position control with real-time feedback, capable of arbitrary user-defined waveforms including multi-frequency superposition of at least three simultaneous components.
Pre-programmed, fixed-routine-only calibrators are not sought. • A motion envelope on the order of 0.1 to 10 Hz in frequency, sub-millimeter to approximately twenty millimeters in displacement, and up to approximately one g in acceleration, with demonstrated small-amplitude positioning accuracy at the sub-millimeter level. • An integrated position sensor and an integrated reference accelerometer, each factory calibrated with NIST-traceable certificates. • Safety features including a hard-wired emergency stop, motion limits, over-current protection, and fault detection with automatic safe shutdown. • Benchtop form factor suitable for standard laboratory power (110 to 120V, or universal input) without special power conditioning or floor reinforcement. • Control software operable in a standalone configuration with no required internet, cloud, or subscription, a programmable interface compatible with MATLAB, Python, or an equivalent industry-standard environment, and non-proprietary data import and export. • Software and electronic content conforming to Section 508 accessibility standards, supported by a Voluntary Product Accessibility Template or equivalent.