The response deadline has passed. Review the details for future reference or to track similar opportunities.
The Department of Veterans Affairs is procuring a liquid handler for the Southern Arizona VA Health Care System. The equipment must feature at least five independent pipetting channels, non-contact dispensing, and integrated camera vision verification. Key technical requirements include a volume capability from 0.5 uL to 20 uL and an external API for system integration.
Provision of a liquid handler system meeting specific pipetting, dispensing, and software integration requirements.
The purpose of this amendment is to modify the salient characteristics for solicitation 36C26226Q0899.
The salient characteristics are hereby modified to read: 5.
Salient Characteristics 5.1 MUST HAVE at least five (5) independent Pipetting channels. 5.2 MUST HAVE Volume capability from 0.5 uL to 20 µL.
With capability to adjust to future workload. 5.3 MUST HAVE Non-contact dispensing to 0.2 µL. 5.4 MUST HAVE Compatible with liquids of varying viscosity and volatility. 5.5 MUST HAVE Liquid level and well bottom sensing. 5.6 MUST HAVE Integrated camera/vision verification. 5.7 MUST HAVE Precision CV < 5% at 0.5 µL. 5.8 MUST HAVE Independent reach for plates, wells, and tubes. 5.9 MUST Reduce consumable waste. 5.10 MUST HAVE User-friendly software and intuitive programming. 5.11 MUST HAVE Real time run monitoring with safety interlocks. 5.12 MUST HAVE External API for system integration.
Submission of your response shall be received not later than Friday, July 3, 2026 at 10:00am PT Local Time.